[1]YANG Li,LIU Feng,GUAN Chaokang,et al.Analysis on Comparison Observation Test Results of Two Different Types of Mercury Detector[J].Journal of Seismological Research,2016,39(03):479-485.
Copy
Journal of Seismological Research[ISSN 1000-0666/CN 53-1062/P] Volume:
39
Number of periods:
2016 03
Page number:
479-485
Column:
Public date:
2016-07-23
- Title:
-
Analysis on Comparison Observation Test Results of Two Different Types of Mercury Detector
- Author(s):
-
YANG Li1; LIU Feng1; GUAN Chaokang2; ZHANG Shuhua2; ZHU Peiyao1
-
(1.Earthquake Administration of Yunnan Province, Kunming 650224, Yunnan, China)(2.Earthquake Administration of Mile Multiple, Mile 652300, Yunnan, China)
-
- Keywords:
-
Hg observation; mercury detector; comparative analysis; co-seismic response
- CLC:
-
P315.63
- DOI:
-
-
- Abstract:
-
Under the same observation condition, we selected Midongshao Well in Mile and compared the observation curves recorded by ATG-6138M Trace Mercury Automatic Analyzer and DFG-B and RG-BQZ Digital Intelligent Mercury Detectors. The comparison result of observation curve and the instrument parameter showed that the detection limit, sensitivity and stability of ATG-6138M Trace Mercury Automatic Analyzer were better than that of DFG-B and RG-BQZ Digital Intelligent Mercury Detectors. The observation curve recorded by ATG-6138M Trace Mercury Automatic Analyzer had the obvious the daily variation, and could record the co-seismic variation.